Conferences \ Circulating Biomarkers World Congress 2015 \ Circulating DNA, Circulating RNA, Circulating Tumor Cells \ Agenda \ Alexander Wolf |
Register | Login |
Purification of Circulating Cell-Free DNA from Plasma using the Automated QIAsymphony SP InstrumentMonday, 23 March 2015 at 14:00 Add to Calendar ▼2015-03-23 14:00:002015-03-23 15:00:00Europe/LondonPurification of Circulating Cell-Free DNA from Plasma using the Automated QIAsymphony SP InstrumentSELECTBIOenquiries@selectbiosciences.com A new ccfDNA extraction protocol on the QIAsymphony® SP revealed a highly efficient recovery of total ccfDNA and fetal ccfDNA compared to the QIAamp® Circulating NA Kit. Linearity was shown for ccfDNA yield from 2-6 ml plasma input. A high efficiency for extraction of low copy numbers of ccfDNA is given. Compatibility of eluates to manifold downstream assays was confirmed for different PCR and methylation-sensitive assays, library preps and subsequent NGS-based analysis of chromosomes. Concluding, the novel protocol enables automated ccfDNA recovery from up to 4 ml plasma and up to 96 samples per QIAsymphony® SP run in 6 hours combined with high recovery efficiency. |